46
Example.2 In case of net value cancelled 300 g of tare weight set in ex.1
The measurement from -300 to 300 g becomes the first range (scale interval 0.1 g).
The measurement from 300 to 600 g becomes the second range (scale interval 0.5 g).
The measurement from 600 to 700 g becomes the second range (scale interval 1 g).
When it is used only within the first range (C function CF-40 : OFF), it is not necessary to set the range.
Please make the relations of the scale interval of the each range as the 1st range
<
the 2nd range
<
the 3rd range.
Please make the relations of the each boundary value of each range as the second range
≦
the third range
≦
the
maximum capacity.
The range upper bound within the range of second is a maximum capacity for the switch by two ranges.
The range upper bound within the range of third is a maximum capacity for the switch by three ranges.
The comparator setting and the fixed tare weight cancellation setting become the scale interval set in the first
range.
6-11. Setting of the stability detection time width in calibration
The stability detection time width in calibration can be set by C function CF-97.
Setting range : [00] ~ [99], Unit : 0.1 s
[05] is set as a default.
6-12. Digital linearizer clear
The compensation data set in the digital linearization can be cleared by the C function CF-98.
If key
is pressed with
display, becomes a blinking display.
Press
key to discontinue the digital linearization at this point.
It becomes a stand-by condition, and the digital linearization is not clear.
is displayed If
key is pressed when displays blinking, and the
operation of digital linearization is completed.
6-13. Memory clear
A clear memory is executed by C function CF-99, and the setting contents of C function is returned to the
default setting.
If
key is pressed when
displays blinking,
becomes a blinking
display. Press
key to discontinue the clear memory at this point.
It becomes a stand-by condition, and digital linearization is not clear.
If
key is pressed when
displays blinking, the display section is turned off and
becomes to
display, then the operation of memory clear is completed.
0 g
300 g
600 g
1 000 g
0 g
-300 g
300 g
600 g 700 g
Scale interval 0.1 g
Scale interval 0.5 g
Scale interval 1 g
Net value
Measurement
value
Gross Value
Tare weight