14
4-4. Explanation for set items before calibration
4-4-1. Flow of set items before calibration
Step 1
Connect the digital load cell with this instrument.
Step 2
To stabilize this unit and the measuring section (load cell),
please make the instrument to the status of energizing for
about 10 min.
Step 3
Register the S/N (serial No.) of digital load cell connecting
the instrument. If already registered, refer to the register
-ing contents. Go to step 4 if unnecessary to refer the S/N.
Step 4
Switchover to the corner adjustment mode.
Go to step 6 if already adjusted or unnecessary to adjust.
Step 5
Execute corner adjustment.
Step 6
Switchover to the calibration mode.
Display the error code [Er-50] right after power up in case of not register the S/N of digital load cell.
Display the error code [Er-50] right after power up if registered S/N (serial No.) of digital load cell has some fault
(X for ID No.)
4-5
Calibration procedures
Connection with digital load cell
Energizing for 10 min after power is
turned on.
4-4-2
Register / refer the S/N to digital load cell
4-4-3
Switchover the corner adjustment mode
4-4-4
Corner adjustment