104
13-3-4. Pin configuration and wiring of terminals
Located the furthest place from the Host (personal computer, sequencer), make the TRM. Terminal and RDB
terminal short, and then connect with the built-in terminal resistance.
For the connection, twisted pair type wire shall be recommended.
The SG of RS-422/485 and the COM.1 of external control input are common.
The internal circuit is isolated from photo coupler.
To suit the instrument to the suitable JIS standard, use the shielded cable, and connect the shielded cable with
the FG terminal.
(1) Connection of 1 to 1
SDA Differential
)
SDB Differential
output(-)
RDA Differential
input(+)
RDB Differential
input(-)
TRM Terminal
resistance
S.G. Signal
grand
RDA
RDB
SDA
SDB
S.G.
SDA
SDB
RDA
RDB Address 00
TRM.
S.G.
SDA
SDB
RDA
RDB Address 00
TRM.
S.G.
RDA
RDB
SDA
SDB
S.G.
RS-422 Host
RS-485 Host
CSD-401
CSD-401
SDA
SDB
RDA
RDB
TRM.
S.G.