13
4-3. Items set after the calibration, if necessary.
(1) Display position of decimal point (C function:
)
The decimal point is put on the load display of this unit.
(2) Unit (C function:
)
The unit is put on the RS-232C and RS-422/ 485 interface load display of this unit.
(3) Digital linearization (
)
It is a function the compensation of three points or less except the zero and span, and to reduce the
measurement error. Refer to the paragraph 4-5. Digital linearization.
(4) Automatic range switch (C function:
~
)
It is the setting of the second range or the third range for [Multi scale interval scale]. The boundary
value and the scale interval within the each range are set. Refer to the 6-10. Automatic range
switch.
(5) Gravity acceleration compensation (C function:
,
)
It is a function for the compensation of the span error by setting the gravity acceleration of the two
different districts with a calibration place and a place to be used. Refer to 6-8. Gravity acceleration
compensation
Please execute the calibration when the applied environment changes if necessary.
The display resolution that the performance becomes effective is 10 000 or less.
The display resolution is a value into which weighting is divided by the scale interval of the first range.
The mass of the weight used in the span calibration must use the one of 2/3 or more of weighing to reduce the
calibration error.