19
Display
.
Put the counterweight for corner adjustment on the ID
No. 5 digital load cell, then press the
key.
Press the
key when the
display is
flickering and the [STAB. mark] lights on.
Memorize each digital load cells data at [P-5] then go
to the next step.
Display
Put the counterweight for corner adjustment on the ID
No. 2 digital load cell, then press the
key. Press
the
key when the
display is
flickering and the [STAB. mark] lights on.
Memorize each digital load cells data at [P-2] then go
to the next step.
Display.
Put the counterweight for corner adjustment on the ID
No. 3 digital load cell, then press the
key.
Press the
key when the
display is
flickering and the [STAB. mark] lights on.
Memorize each digital load cells data at [P-3] then go
to the next step.
Display.
Put the counterweight for corner adjustment on the ID
No. 4 digital load cell, then press the
key.
Press the
key when the
display is
flickering and the [STAB. mark] lights on.
Memorize each digital load cells data at [P-4] then go
to the next step.
Display
after finish corner adjustment.
Press
key for finish corner adjustment.
Display
and memorize the setting data
in internal memory.
Press
key to stand-by condition.
Complete the corner adjustment.
Use a counterweight for corner adjustment with minimum 1/20 of the measuring weight to maximum the
measuring weight.
Stand-by condition
[A]
No next ID
Press key
after STAB
mark lights on.
Press key
after STAB
mark lights on.
Press key
after STAB
mark lights on.
Press key
after STAB
mark lights on.
No next ID
No next ID
No next ID
[B]
[C]
Summary of Contents for CSD-401