125
17. Error display
Flashing displays about 2 seconds when selecting No. whose No. is not existed on
the list at the time of setting C function and function is set.
During calibration is made, flashing displays about 2 seconds, at the time of setting
[Weighing capacity (DISP) < setting weight value (LOAD)].
EEPROM writing error. Contact with us.
EEPROM reading error. Contact with us.
During calibration, flashing displays in about 2 seconds when the output count of
digital load cell is less than ‘- 1 100 000× connecting number of digital load cell’ and
at the same time more than the Zero adjustment range at minus side.
During calibration, flashing displays in about 2 seconds when the output count of
digital load cell is more than [1] 100 000× connecting number of digital load cells’ and
at the same time more than the Zero adjustment range at pulse side.
During calibration, flashing displays in about 2 seconds when [Output count of digital
load cell of Span] – [Output count of digital load cell at zero point] <0 and also before
reaching the span adjustment range.
During calibration, flashing displays in about 2 seconds when ‘output count of digital
load cell of span’ - ‘output count of digital load cell of zero point’ > 1 100 000×number
of connection of digital load cell, and also exceeding the span adjustment range.
At the time of digital liberalized compensation is made, flashing displays in about 2
seconds when the output of digital load cell is low.
At the time of digital liberalized compensation is made, flashing displays in about 2
Seconds when the output of digital load cell is high.
(Effective for the
C
function CF-15 ‘Power on zero operation’)
Displays when power is supplied at the time of output of digital load cell exceeds
±10 % of weighing capacity, or when the display is ON.
Pressing the
key with the condition of
, the load value display shall be
appeared compulsively.
Lights on when the load display exceeds ‘(+ weighing capacity) + 9D’, or exceeds
‘+110 % of weighing capacity’. (Due to the setting of
C
function CF-03.)
Lights on when the load display exceeds ‘(- weighing capacity ) - 9D’ or exceeds ‘-
20D’. (Due to the setting of
C
function CF-03.)
Output of pulse over of digital load cell
(Even the output count of digital load cell of one set is 1 100 000 or more.)
Output of minus over of digital load cell
(Even the output count of digital load cell of one set is less then -1 100 000.)
All lights on.
At the time of feeding power, or display ON, you can’t detect the stability.
When you make the stability detection OFF with whichever Function F-10 and
Function F-11, then all lights on.
When S/N (serial No.) of digital load cell isn’t registered for the C Function from
CF-60 to CF-67, displays soon after power supply is on.
When the S/N (serial no.) of digital load cell of ID=1 and the S/N set for the C
Function CF-60 is different, displays soon after power supply is on.
When the S/N (serial no.) of digital load cell of ID=2 and the S/N set for the C
Function CF-61 is different, displays soon after power supply is on.
When the S/N (serial no.) of digital load cell of ID=3 and the S/N set for the C
Function CF-62 is different, displays soon after power supply is on.