18
4-4-4. Corner adjustment
From
display to press
key, it
displays
At this point, do not apply any load in metering section
without tare weight.
Pressing
key and displays
is
flickering then start the zero point calibration.
After [STAB. mark] lights on, press
key.
Cancel the corner adjustment pressing
key and
displays
Complete zero point calibration and displays
Put the counterweight for corner adjustment on the ID
No. 1 digital load cell, then press the
key.
Press the
key when the
display is
flickering and the [STAB. mark] lights on. Memorize
each digital load cells data at [P-1] then go to the next
step.
Display
.
Put the counterweight for corner adjustment on the ID
No. 2 digital load cell, then press the
key. Press
the
key when the
display is
flickering and the [STAB. mark] lights on.
Memorize each digital load cells data at [P-2] then go
to the next step.
Display
.
Put the counterweight for corner adjustment on the ID
No. 3 digital load cell, then press the
key.
Press the
key when the
display is
flickering and the [STAB. mark] lights on.
Memorize each digital load cells data at [P-3] then go
to the next step.
Display
.
Put the counterweight for corner adjustment on the ID
No. 4 digital load cell, then press the
key.
Press the
key when the
display is
flickering and the [STAB. mark] lights on.
Memorize each digital load cells data at [P-4] then go
to the next step.
Press key after
STAB mark lights
[A]
[B]
Press key
after STAB
mark lights on.
Press key
after STAB
mark lights on.
Press key
after STAB
mark lights on.
[C]
No next ID
No next ID
No next ID