4-8-2. Calibration mode only of zero point
Please make the display to
by pressing
key from the display of
.
Please make to the status where nothing is put on the
measuring section here excluding the tare outside maximum
capacity.
The display of
blinks by pressing
key,
and start the calibration at zero.
Press
key after the stability mark lighting on.
The calibration of zero point is completed, and
is displayed.
Press
key to finish the calibration mode only of zero
point.
is displayed, and the setting data is memorized
in the internal memory.
Please set to the stand-by condition by pressing
key.
The calibration mode only of zero point is finished key.
Error display of the calibration at zero
: The display blinks for about 2 seconds when the digital load cell output count is
less than -1100 000
×
number of digital load cell connecting.
(Refer to the paragraph 17. Error display.)
: The display blinks for about 2 seconds when the load cell output is over -1100
00
×
number of digital load cell connecting.
(Refer to the paragraph 17. Error display.)
Stand-by condition