4-5-7. Calibration at span
When span calibration starts,
is displayed. Please select span calibration method.
(1) Method by the measuring data
Press
key, and execute the operation of a).
Execute the span calibration with weight put on the measuring section.
(2) Method by the numeric output of the digital load cell output count
Press
key, and execute the operation b).
Execute the span calibration by inputting the value which subtract the output count of the digital
load cell to zero point of measurement set in paragraph 4-5-6. from the digital load cell output count
corresponding to the mass of the counterweight set in paragraph 4-5-5.
a) Method with weight
Put the counterweight corresponds to mass set in the paragraph
4-5-5 on the measuring section.
Press
key when
display blinks, and STAB.
mark lights on. Span is memorized and
is
displayed.
b) Method by the numeric input of digital load cell output count.
When
key is pressed,
is displayed, and when
key is pressed,
is displayed also.
[
] is memorized output count of a digital load cell. Set
the value in count scale which subtract the output count of the
digital load cell to zero point of measurement from the output
count of the digital load cell corresponding to the mass of the
counterweight.
[
] is set the lower four count value.
[
] is set the higher three count value.
: The value of changed digit is changed.
: Changed digit is selected.
: The displayed value is changed to [0].
: Setting is interrupted and display returns to
: Display value is memorized, and it proceed to next step.
After the setting, press
key.
《
Supplementary explanation
》
Flickering point moves to the display
first digit when
press
key on the fourth digit of
display.
Error display in the calibration of zero point
The display is flickering for about two seconds when ([Digital load cell output
count at span] – [Digital load cell output count at zero point])
≦
0. (Refer to the
paragraph 17. Error display) (Refer to the paragraph 17. Error display.)
The display is flickering for about two seconds when ([Digital load cell output
count at span] – [Digital load cell output count at zero point]) 1 100 000
×
number of connecting digital load cell. (Refer to the paragraph 17. Error display.)
Press
key with
weight on the
measuring
section.
Output count of
digital load cell
is memorized.
Count
Count
Output count set
for digital load cell.
Press
key after
STAB mark
lights on.