45
6-10. Automatic range switch
Automatic range switch function switches automatically and displays the scale interval by the measuring
data of gross or net value by dividing the range of the measurement to the maximum capacity into 2 or 3.
6-10-1. Setting of the range switch operation
The range switch operation can be selected by C function CF-40.
The range switch can be selected from [OFF (without range switch)], [Two range switch] and [Three
range switch]. [OFF] is selected as a default.
6-10-2. Setting the scale interval of the second range
The scale interval of the second range can be set by the C function CF-41.
The scale interval can be selected from [1 scale interval], [2 scale intervals], [5 scale intervals], [10 scale
intervals], [20 scale intervals] and [50 scale intervals]. [5 scale intervals] is selected as a default.
6-10-3. Setting the boundary value of the second range
The boundary value of the second range can be set by C function CF-42.
Setting range : [0] ~ [999999]. [003000] is selected as a default
6-10-4. Setting the scale interval of the third range
The scale interval of the third range can be set by the C function CF-43.
The scale interval can be selected from [1 scale interval], [2 scale intervals], [5 scale intervals], [10 scale
intervals], [20 scale intervals] and [50 scale intervals]. [10 scale intervals] is selected as a default.
6-10-5. Setting the boundary value of the third range
The boundary value of the third range can be set by C function CF-44.
Setting range: [0] ~ [999999]. [006000] is selected as a default
ex.1 In case of gross value
The first range: 300.0 g (Boundary of the second range), scale interval 0.1 g (Scale interval in calibration)
The second range: 600.0 g (Boundary of the third range), scale interval 0.5 g (Scale interval of the
second range)
The third range : 1 000.0 g (The maximum capacity), scale interval 1 g (Scale interval of the third range)
The measurement from 0 to 300 g becomes the first range (scale interval 0.1 g).
The measurement from 300 to 600 g becomes the second range (scale interval 0.5 g).
The measurement from 600 to 1 000 g becomes the second range (scale interval 1 g).
0 g
300 g
600 g
1 000 g
Scale interval 0.1 g
Scale interval 0.5 g
Scale interval 1 g
Measurement
value