6-30
3. Unscrew the two M3 nuts and remove the battery block plate to replace the battery, as
shown below.
6.2.27 Remove the Power Cord
Unscrew the three screws as shown below to remove the power cord.
WARNING
z
Before replacing the power cord, be sure to disconnect the power supply to
prevent electric shock.
6.2.28 Remove the Isolation Transformer
1. Remove the two drawer assemblies.
2. Remove the trolly rear panel assembly.
Nut