4-56
4.4.4 Load Gas Module or BIS Module
The anesthesia machine provides online upgrade of gas modules and BIS module.
4.4.4.1 Check the Corresponding Relationship between the Loaded
Module and its Software Function
The following table lists the modules which can be loaded and their respective functions.
Module name
Functional description
Loading name of module
software function
AG
module
Provides AG and CO2 concentration
monitoring and O2 concentration monitoring
as well if O2 module is configured.
AG
M02B CO2
module
Provides CO2 concentration monitoring
CO2
Mainstream CO2
module
Provides CO2 concentration monitoring
CO2
Microstream
CO2 module
Provides CO2 concentration monitoring
CO2
BIS module
Provides BIS monitoring
BIS
4.4.4.2 How to Load the Software Function of Gas Modules or BIS Module
To load the software function of the gas modules or BIS module:
1. Select
the
[
Maintenance
] shortcut key
→
[
Factory Maintenance >>
]
→
enter the
required password
→
[
Select Module >>
] to open the module selection menu. Select
one or more from [
CO2
], [
AG
] and [
BIS
].
2. After the selection, restart the anesthesia machine or re-plug in the module to activate
the added module.
Summary of Contents for WATO EX-55
Page 1: ...WATO EX 55 65 Anesthesia Machine Service Manual ...
Page 2: ......
Page 4: ...II FOR YOUR NOTES ...
Page 6: ...IV FOR YOUR NOTES ...
Page 12: ...6 FOR YOUR NOTES ...
Page 58: ...3 22 FOR YOUR NOTES ...
Page 242: ...6 54 6 4 1 Electrical Connections Electrical Circuit Diagram ...
Page 260: ...6 72 FOR YOUR NOTES ...
Page 261: ...7 1 7 Parts 7 1 Main Unit 7 1 1 Main Unit 7 1 1 1 Exploded View ...
Page 321: ......
Page 322: ...P N 0621 20 78654 1 1 ...