2-20
or negative pressure occurs inside the gas reservoir, gas is inputted or outputted to ensure
pressure balance inside the system.
The AGSS transfer system is a blue tube with 30 mm conical connectors at both ends. The
inlet of the transfer system is a female 30 mm conical connector and the outlet a male 30 mm
conical connector. The transfer system is connected to the receiving system through the male
30 mm conical connector. The receiving system is connected to the receiving hose through
the proprietary connector. The receiving hose is connected to the hospital’s disposal system
through BS 6834 connector. The following picture shows the structures of and the
connections between the AGSS transfer system, receiving system and disposal system.
Receiving system
Receiving hose
To the disposal system
Hook
Transfer system
To the exhaust port