6-1
6
Repair and Disassembly
WARNING
z
To help prevent fires, only use lubricants approved for anesthesia or O
2
equipment.
z
Do not use lubricants that contain oil or grease. They burn or explode in high O
2
concentrations.
z
Obey infection control and safety procedures. Used equipment may contain blood
and body fluids.
z
Movable part and removable components may present a pinch or a crush hazard.
Use care when moving or replacing system parts and components.
z
Use care when disassembling the parts with sharp edges to avoid cuts.
z
Pay attention to the screws during the disassembly to prevent screws from falling
into the inside of the equipment. Failure to do so may cause short circuit.
z
Make sure to bleed gas pressure before disassembling pneumatic fittings to avoid
personal injury caused by high pressure gas.
NOTE
z
When re-assembling, inspect all parts for deterioration. Replace them if
necessary. Use appropriate screws and parts.
z
After repairs are completed or parts replaced, perform the checkout procedure.
Refer to3Checkout and Test.