4-48
Before upgrading the module software, note the matching between module name and module
code as listed below.
Module name
Module code
Monitor module
MC
Auxiliary control module
MC2
Power board
POWER
Electronic flowmeter
FLOW
MO2B CO2
0611-MO2B
Main control board FPGA
display drive
FPGA
To upgrade the module software:
1. When selecting the module software upgrade package, confirm the correctness of
checksum and version. You also need to check the module code, as shown below.
2. Other operations are similar to those for booting software upgrade. Refer to
4.4.1.2Booting Software Upgrade
to finish the upgrade.
NOTE
z
After completing module software upgrade, turn on the anesthesia machine to
confirm the correctness of upgrade software version information.