5-36
6. Release the negative pressure ball. If the negative pressure ball is fully expanded
within 30s during one of the two tests, it indicates that the system switch assembly is
damaged. Handle this problem as described in the troubleshooting table.
7. Turn on the system switch.
8. Occlude the pulled-out tube end on the system switch assembly by using 3126-06-00
tube plug.
9. Flatten the negative pressure ball to remove the gas inside. Then re-install the gas
outlet switch of the negative pressure ball properly. Connect the other end of the
negative pressure ball to the pulled-out end of No.45 or 51 PU tube in turn through
3106-06-00 adapter connector, as shown below.