3-1
3
Checkout and Test
WARNING
z
After servicing the equipment or replacing its components, complete all the tests in
this section.
z
Before doing the tests in this section, completely reassemble the equipment and
refer to 4Maintenance and Calibration to do necessary calibrations.
3.1 System Inspection
NOTE
z
Make sure that the breathing circuit is correctly connected and not damaged.
z
The top shelf weight limit is 30 kg.
WARNING
z
Do not leave gas cylinder valves open if the pipeline supply is in use. Cylinder
supplies could be depleted, leaving an insufficient reserve supply in case of pipeline
failure.
Make sure that:
1. The equipment is not damaged.
2. All components are correctly attached.
3. The breathing circuit is correctly connected and the breathing tubes are not damaged.
4. The vaporizers are locked in position.
5. The gas supplies are connected and the pressures are correct.
6. Cylinder valves are closed on models with cylinder supplies.
7. The casters are not loose and the brake (s) is set and prevents movement.
8. Make sure the circuit is locked safely (in the
position).
9. The power cord is correctly connected. The AC mains indicator and the battery indicator
work normally.
10. The anesthesia machine is switched on or off normally.
Summary of Contents for WATO EX-55
Page 1: ...WATO EX 55 65 Anesthesia Machine Service Manual ...
Page 2: ......
Page 4: ...II FOR YOUR NOTES ...
Page 6: ...IV FOR YOUR NOTES ...
Page 12: ...6 FOR YOUR NOTES ...
Page 58: ...3 22 FOR YOUR NOTES ...
Page 242: ...6 54 6 4 1 Electrical Connections Electrical Circuit Diagram ...
Page 260: ...6 72 FOR YOUR NOTES ...
Page 261: ...7 1 7 Parts 7 1 Main Unit 7 1 1 Main Unit 7 1 1 1 Exploded View ...
Page 321: ......
Page 322: ...P N 0621 20 78654 1 1 ...