5-55
7. Check the APL valve, support tube on the median plate, and sealing components of the
bag/mechanical ventilation switch assembly
The test requires a T-shaped Allen wrench.
Test procedures:
(
1
)
Turn off the system switch.
(
2
)
Remove the APL valve. Check all seals and replace the defective ones.
(
3
)
Remove the support tube on the median plate. Check the seals and replace the
defective one.
(
4
)
Remove the bag/mechanical ventilation switch. Check the seals and replace the
defective one.
8. Check the ACGO assembly
Tools required:
Negative pressure ball (quantity: 1)
Circuit adapter test fixture (quantity: 1)
PU tube (6X100) (quantity: 1)
3106-08-10 adapter connector (quantity: 2)
3126-06-00 tube plug (quantity: 1)
3126-08-00 tube plug (quantity: 1)
3126-10-00 tube plug (quantity: 2)
Test procedures:
(
1
)
Turn off the system switch.
(
2
)
Pull out No.21 and 22 PU tubes which connect the ACGO assembly to the circuit
adapter assembly. The ends of the tubes which connect the ACGO assembly are pulled
out but the other ends are not, as shown below.
Seal to be checked