4-22
4. Connect the anesthesia machine calibration device to the power source. The following
pictures show the connectors on the calibration device.
5. Start the calibration device to enter the startup screen followed by sensor heating screen
(waiting for approximately 5 minutes as required by the prompt message) and then
zeroing screen. Press the “ZERO” key on the panel. After the new screen pops up, press
the “OK” key on the panel to complete zeroing.
NOTE
z
When zeroing the anesthesia machine calibration device, make sure that no gas
flows through the device, or unplug the tube connected to the gas inlet of the
device.
6. Connect the calibration communication connector of the calibration device to that of the
anesthesia machine by using the special communication cable. The calibration device
can communicate with the anesthesia machine through two types of connection:
Calibration communication
connector
Power connector
High flow inlet
(0 to 120 L/min)
High flow outlet
(0 to 120 L/min)