5.3.1.3 Precautions for Use of Vaporizer Manifold Test Fixture
When using the vaporizer manifold test fixture, remove the seal between the connector of
vaporizer manifold assembly and the vaporizer first. Then slide the test fixture into the
connector, as shown below.
Turn the pressure head of vaporizer manifold test fixture clockwise until the bottom surface
of the pressure head is in contact with the top surface of the connector of the vaporizer
manifold assembly, as shown below.
After the test fixture is slid in
When the test fixture is slid in
Contact here when mounted in position