5-32
6. Flatten the negative pressure ball to remove the gas inside. Then re-install the gas
outlet switch of the negative pressure ball properly. Connect the other end of the
negative pressure ball to the pulled-out end of No.25 PU tube through 3106-06-00
adapter connector, as shown below.
7. Release the negative pressure ball. If the negative pressure ball is fully expanded
within 30s, it indicates that the float flowmeter related assembly is damaged. In this
case, perform the following operations.
8. Remove the left front panel.
9. Pull out No.26 PU tube which connects the throttling device to the float flowmeter.
The end of the tube which connects the float flowmeter is pulled out but the other end
is not.
10. Occlude the pulled-out tube end on the float flowmeter by using 3126-06-00 tube plug.
11. Flatten the negative pressure ball to remove the gas inside. Then re-install the gas
outlet switch of the negative pressure ball properly. Connect the other end of the
negative pressure ball to the pulled-out end of No.25 PU tube through 3106-06-00
adapter connector, as shown below.