3-16
3.9.2 Test the O
2
Concentration Monitoring and Alarms
NOTE
z
This test is not required if O
2
sensor is not configured.
1. Set the bag/mechanical ventilation switch to the
position.
2. Remove the O
2
sensor and make sure that the sensor measures approximately 21% O
2
in
room air.
3. Select
the
[
Alarm Setup
] shortcut key and then [
Ventilator >>
]. Set the FiO
2
low alarm
limit to 50%.
4. Make sure that a low FiO
2
alarm occurs.
5. Set the FiO
2
low alarm limit to a value less than the measured FiO
2
value and make sure
that the alarm cancels.
6. Put the O
2
sensor back in the circuit.
7. Select
the
[
Alarm Setup
] shortcut key and then [
Ventilator >>
]. Set the FiO
2
high
alarm limit to 50%.
8. Connect the manual bag to the manual bag port. Push the O
2
flush button
to fill the
manual bag. After two to three minutes, make sure that the sensor measures
approximately 100% O
2.
9. Make sure that a high FiO
2
alarm occurs.
10. Set the FiO
2
high alarm limit to 100% and make sure that the alarm cancels.
3.9.3 Test the Low Minute Volume (MV) Alarm
1. Make sure that MV alarm is turned on.
2. Select
the
[
Alarm Setup
] shortcut key and then [
Ventilator >>
]. Set the MV low alarm
limit to 8.0 L/min.
3. Make sure that a low MV alarm occurs.
4. Select
the
[
Alarm Setup
] shortcut key and then [
Ventilator >>
]. Set the MV low alarm
limit to the default.
Summary of Contents for WATO EX-55
Page 1: ...WATO EX 55 65 Anesthesia Machine Service Manual ...
Page 2: ......
Page 4: ...II FOR YOUR NOTES ...
Page 6: ...IV FOR YOUR NOTES ...
Page 12: ...6 FOR YOUR NOTES ...
Page 58: ...3 22 FOR YOUR NOTES ...
Page 242: ...6 54 6 4 1 Electrical Connections Electrical Circuit Diagram ...
Page 260: ...6 72 FOR YOUR NOTES ...
Page 261: ...7 1 7 Parts 7 1 Main Unit 7 1 1 Main Unit 7 1 1 1 Exploded View ...
Page 321: ......
Page 322: ...P N 0621 20 78654 1 1 ...