3-20
3.11 Power Failure Test
1. Connect the anesthesia machine to the AC power source. Both AC power LED and
battery LED should come on. If the AC power LED is not lit, check the fuse and power
board.
2. Set the system switch to the
position.
3. Unplug the power cord with the system turned on. The message [
Battery in Use
] is
displayed. Meanwhile, the AC power LED is extinguished and the battery LED is
flashing.
4. Reconnect the AC power. The prompt message disappears. The AC power LED is
illuminated. The battery LED stops flashing and stays ON.
Nylon pad
Connector of the disposal
system
Pressing plate
Receiving hose
O-ring
Receiving hose
Pressing plate
Connector of the
receiving system
Battery LED
AC power LED Operating state LED
Summary of Contents for WATO EX-55
Page 1: ...WATO EX 55 65 Anesthesia Machine Service Manual ...
Page 2: ......
Page 4: ...II FOR YOUR NOTES ...
Page 6: ...IV FOR YOUR NOTES ...
Page 12: ...6 FOR YOUR NOTES ...
Page 58: ...3 22 FOR YOUR NOTES ...
Page 242: ...6 54 6 4 1 Electrical Connections Electrical Circuit Diagram ...
Page 260: ...6 72 FOR YOUR NOTES ...
Page 261: ...7 1 7 Parts 7 1 Main Unit 7 1 1 Main Unit 7 1 1 1 Exploded View ...
Page 321: ......
Page 322: ...P N 0621 20 78654 1 1 ...