6-10
3. Turn over the assembly. Disconnect the connection lines to the inverter and unscrew
the four screws to remove the display, as shown below.
6.2.8 Remove the Flowmeter Control Board
1. Follow the steps described in
6.2.7Disassemble the Display Assembly
2. Unscrew the four screws to remove the electronic flowmeter control board, as shown
below.
6.2.9 Remove the Total Flowmeter
1. Follow the steps described in
6.2.5Remove the Left Front Panel Assembly.