4-45
2. Click
[
Select Package
] to enter the [
Select Package
] menu.
4. Click
[
>>>
]. Select the booting software of the anesthesia machine (Code: BIOS) in the
dialog box and then click [
Open
].
5. Check the checksum and version of the booting software as shown below.
6. If the checksum and version are correct, click [
Ok
].
7. Click
[
Start (Single)
] on the main screen.
8. Re-start the anesthesia machine to start to upgrade the booting software.
9. Wait for the message prompting upgrade success. Select the [
Maintenance
] shortcut
key
→
[
Factory Maintenance >>
]
→
[
System Information >>
] to check the version
information of the booting software.
NOTE
z
After completing booting software upgrade, turn on the anesthesia machine to
confirm the correctness of booting software version information.