4-50
4.4.2 Software Function Activation
For system software version greater than V03.01.00, online upgrade is supported. The factory
can activate all the functions listed in the following table through activation codes. When the
user wants to add any function listed in this table, the service engineer can apply to the
factory for activation code so as to activate the corresponding function.
Function Description
PCV
Pressure control ventilation mode
PSV
Pressure support ventilation mode
SIMV Synchronized
intermittent
mandatory ventilation mode
Spirometry loops
F-V and P-V loops
4.4.2.1 Apply for Software Function Activation Code
NOTE
z
To apply for activation code, you must provide the relevant information of the
anesthesia machine whose configurations are to be activated, such as machine ID,
existing configuration and configuration to be activated.
When the user wants to add any paid configuration listed in the above table, the service
engineer must apply to the Service Department for the software function activation code
based on the user’s need.
1. Record the serial number of the anesthesia machine (see the label on the left side of the
machine) whose configurations are to be activated.
2. Record the machine ID and the current configuration of the anesthesia machine. Select
the [
Maintenance
] shortcut key
→
[
User Maintenance >>
]
→
[
Configuration
Information >>
] to open the configuration information menu as shown below.
Summary of Contents for WATO EX-55
Page 1: ...WATO EX 55 65 Anesthesia Machine Service Manual ...
Page 2: ......
Page 4: ...II FOR YOUR NOTES ...
Page 6: ...IV FOR YOUR NOTES ...
Page 12: ...6 FOR YOUR NOTES ...
Page 58: ...3 22 FOR YOUR NOTES ...
Page 242: ...6 54 6 4 1 Electrical Connections Electrical Circuit Diagram ...
Page 260: ...6 72 FOR YOUR NOTES ...
Page 261: ...7 1 7 Parts 7 1 Main Unit 7 1 1 Main Unit 7 1 1 1 Exploded View ...
Page 321: ......
Page 322: ...P N 0621 20 78654 1 1 ...