4-55
3. Select
[
Ok
] to confirm the selection. If flowmeter standard or pipeline is changed, the
message [
Please restart the anesthesia machine to activate the selection.
] is
prompted as shown below.
4. Select
[
Ok
] and then restart the anesthesia machine to activate the selected
configuration.
5. Check and confirm that the software configuration of electronic flowmeters is changed
as expected.
NOTE
z
Powering off or restarting the anesthesia machine before the message [Please
restart the anesthesia machine to activate the selection.] is prompted can damage
the BIOS program on the main control board.
z
After the software configuration of electronic flowmeters is changed, restart the
anesthesia machine to activate the configuration.
4.4.3.3 Commonly-encountered Problems and Recommended Actions
Failure description
Possible cause
Recommended action
After changing the software
configuration of electronic
flowmeters, restart the anesthesia
machine. The main control board
buzzes long, resulting in restart
failure.
The anesthesia machine is turned
off or powered off before
software configuration change
success is prompted, which
damages the BIOS program on
the main control board.
Return the main control
board to factory for
repair.
After changing the software
configuration of electronic
flowmeters, the electronic
flowmeters on the anesthesia
machine are inconsistent with
those displayed on the screen in
terms of color and number of
tubes.
The number of tubes and
applicable standard are selected
improperly.
Select correct number of
tubes and applicable
standard for electronic
flowmeters again.
Summary of Contents for WATO EX-55
Page 1: ...WATO EX 55 65 Anesthesia Machine Service Manual ...
Page 2: ......
Page 4: ...II FOR YOUR NOTES ...
Page 6: ...IV FOR YOUR NOTES ...
Page 12: ...6 FOR YOUR NOTES ...
Page 58: ...3 22 FOR YOUR NOTES ...
Page 242: ...6 54 6 4 1 Electrical Connections Electrical Circuit Diagram ...
Page 260: ...6 72 FOR YOUR NOTES ...
Page 261: ...7 1 7 Parts 7 1 Main Unit 7 1 1 Main Unit 7 1 1 1 Exploded View ...
Page 321: ......
Page 322: ...P N 0621 20 78654 1 1 ...