6-71
Tubing
S/N From
To
P/N
1
Differential pressure gauge
connector 1
Monitor board 6
2
Differential pressure gauge
connector2
Monitor board 7
3
Differential pressure gauge
connector 3
Monitor board 5
4
Three-way valve upper
connector 1
Monitor board 1
5
Three-way valve upper
connector 2
Y1-1
6
Three-way valve upper
connector 3
Monitor board 3
7
Three-way valve upper
connector 4
Monitor board 4
8
Y1-2
Monitor board 2
9
Y1-3
Monitor board 9
10
Three-way valve lower
connector 1
Y2-1
A21-000007---
11
Y2-2
Pressure sampling connector
M6G-020046---
12
Three-way valve lower
connector 2
Y2-1 A21-000007---
13
Y2-2
Pressure sampling connector
M6G-020046---
14
Three-way valve lower
connector 3
Y2-1 A21-000007---
15
Y2-2
Pressure sampling connector
M6G-020046---
16
Three-way valve lower
connector 4
Y2-1 A21-000007---
17
Y2-2
Pressure sampling connector
M6G-020046---
18
Gas-out connector of expiratory
valve
Gas-in connector of gas reservoir
M6G-020018---
19
Gas reservoir
Modular rack connector
M90-000049---
21
Gas-out connector of ACGO
assembly
Circuit adapter
22
Gas-out connector of ACGO
assembly
Fresh gas
M6G-020038---
23
Drive gas connector
Drive gas bent pipe
24
APL gas release connector
APL gas release bent pipe
M6G-020018---
Y1 Three-way
connector
M90-100030---
Y2 Two-way
connector
M02A-10-25942
Summary of Contents for WATO EX-55
Page 1: ...WATO EX 55 65 Anesthesia Machine Service Manual ...
Page 2: ......
Page 4: ...II FOR YOUR NOTES ...
Page 6: ...IV FOR YOUR NOTES ...
Page 12: ...6 FOR YOUR NOTES ...
Page 58: ...3 22 FOR YOUR NOTES ...
Page 242: ...6 54 6 4 1 Electrical Connections Electrical Circuit Diagram ...
Page 260: ...6 72 FOR YOUR NOTES ...
Page 261: ...7 1 7 Parts 7 1 Main Unit 7 1 1 Main Unit 7 1 1 1 Exploded View ...
Page 321: ......
Page 322: ...P N 0621 20 78654 1 1 ...