MI 3102(H) BT
Eurotest
XE
Technical specifications
101
10.12
First fault leakage current – ISFL (MI 3102 BT only)
Measuring range (mA)
Resolution (mA)
Accuracy
0.0
19.9
0.1
±(5 % of r 3 digits)
Measuring resistance .................................. approx. 390
Nominal voltage ranges .............................. 93 V
U
L1-L2
134 V
185 V
U
L1-L2
266 V
10.13
Calibrated resistance for IMD testing (MI 3102 BT only)
Threshold indicative insulation resistance
Measuring range (kΩ)
Resolution (
kΩ)
Notes
5
640
5
Indicative values
Up to 128 steps
Nominal voltage ranges .............................. 93 V
U
L1-L2
134 V
185 V
U
L1-L2
266 V
First fault leakage current at threshold insulation resistance
Measuring range (mA)
Resolution (mA)
Note
0.0
19.9
0.1
calculated value*
)
*
)
5.14 Testing of insulation monitoring devices
for more
information about calculation of first fault leakage current at threshold insulation resistance.