MI 3102(H) BT
Eurotest
XE
Technical specifications
99
10.10
TRMS Clamp current
Instrument
Maximum voltage on C1 measuring input ..................................... 3 V
Nominal frequency ................................. 0 Hz, 40 Hz
500 Hz
AC current clamp A1018
Range = 20 A
Measuring range (A)
Resolution (A)
Accuracy*
0.0 m
99.9 m
0.1 m
(5 % of r 5 digits)
100 m
999 m
1 m
(3 % of r 3 digits)
1.00
19.99
0.01
(3 % of reading)
AC current clamp A1019
Range = 20 A
Measuring range (A)
Resolution (A)
Accuracy*
0.0 m
99.9 m
0.1 m
indicative
100 m
999 m
1 m
(5 % of reading )
1.00
19.99
0.01
(3 % of reading)
AC / DC current clamp A1391
Range = 40 A
Measuring range (A)
Resolution (A)
Accuracy*
0.00
1.99
0.01
(3 % of r 3 digits)
2.00
19.99
0.01
(3 % of reading)
20.0
39.9
0.1
(3 % of reading)
Range = 300 A
Measuring range (A)
Resolution (A)
Accuracy*
0.00
19.99
0.01
indicative
20.0
39.9
0.1
40.0
299.9
0.1
(3 % of r 5 digits)
* Accuracy at operating conditions for instrument and current clamp is given.