MI 3102(H) BT
Eurotest
XE
Measurements
73
Illumination measurement procedure
Select the
SENSOR
function from
OTHERS menu
.
Select test
parameters
(optional).
Connect
illumination sensor to the instrument.
Take the position of LUXmeter (see
). Make sure that LUXmeter is turned on.
Press the
TEST
key to start the continuous measurement.
Press the
TEST
key again to stop the measurement.
Store
the result by pressing the
MEM
key (optional).
Figure 5.65: Example of illumination measurement result
Displayed results:
E
............... Illumination
Notes:
For accurate measurement make sure that the milk glass bulb is lit without any shadows
cast by hand, body or other unwanted objects.
It is very important to know that the artificial light sources reach full power of operation
after a period of time (see technical data for light sources) and should be therefore
switched on for this period of time before the measurements are taken.