MI 3102(H) BT
Eurotest
XE
Appendix D
– Commanders
114
PE terminal test procedure
Connect
commander to the instrument.
Connect
commander to the item to be tested (see
Figure D.4
).
Touch PE test probe (the
TEST
key) on commander for at least one second.
If PE terminal is connected to phase voltage both LEDs will light yellow, the warning
message on the instrument
is displayed, instrument’s buzzer is activated, and further
measurements are disabled in Zloop and RCD functions.
Figure D.4: Reversed L and PE conductors (application of plug commander)