MI 3102(H) BT
Eurotest
XE
Measurements
53
Note:
High fluctuations of mains voltage can influence the measurement results (the noise sign
is displayed in the message field). In this case it is recommended to repeat few
measurements to check if the readings are stable.
5.7.2 Voltage drop
The voltage drop is calculated based on the difference of line impedance at connection points
(sockets) and the line impedance at the reference point (usually the impedance at the
switchboard).
Circuits for measurement of voltage drop
Figure 5.33: Phase-neutral or phase-phase voltage drop measurement
– connection of plug
commander and 3-wire test lead
Voltage drop measurement procedure
Step 1:
Measuring the impedance Zref at origin
Select the
Z LINE
function using the function selector keys.
Set sub-function to
ΔU
using
UP / DOWN
keys.
Select test
parameters
(optional).
Connect
test cable to the instrument.
Connect
the test leads to the origin of electrical installation (see
Press the
CAL
key to perform the measurement.
Step 2:
Measuring the voltage drop
Set sub-function to
ΔU
using
UP / DOWN
keys.
Select test
parameters
(Fuse type must be selected).
Connect
test cable or plug commander to the instrument.
Connect
the test leads to the tested points (see
Press the
TEST
key to perform the measurement.
Store the result by pressing the
MEM
key (optional).
Step 1 - Zref
Step 2 - Voltage drop
Figure 5.34: Examples of voltage drop measurement result