Instrument operation
29
4.4.7 Commander support
The support for remote commanders can be
enabled or disabled in this menu.
Figure 4.11: Selection of commander support
Keys:
UP / DOWN
Selects commander option.
TEST
Confirms selected option.
ESC
Exits back to settings menu.
Function
selector
Exits back to main menu without
changes.
Note:
Commander disabled
option is intended to disable the commander’s remote keys. In the
case of high EM interfering noise the operation of the commander can be irregular.
4.4.8 Initial settings
In this menu the instrument settings, measurement
parameters and limits can be set to initial (factory)
values.
Internal Bluetooth module is initialized.
Figure 4.12: Initial settings dialogue
Keys:
UP / DOWN
Selects option [YES, NO].
TEST
Restores default settings (if YES is
selected).
ESC
Exits back to settings menu.
Function
selector
Exits back to main menu without
changes.
Warnings:
Customized settings will be lost when this option is used!
If the batteries are removed for more than 1 minute the custom made settings will be
lost.
The default setup is listed below:
Instrument setting
Default value
Language
English
Contrast
As defined and stored by adjustment procedure
Earthing system*
TN/TT
Lenght units
m
Isc factor
1.00
RCD standards
EN 61008 / EN 61009
Commander
A 1314, A 1401
Internal bluetooth
Initialization of internal Bluetooth module