MI 3102(H) BT
Eurotest
XE
Technical specifications
98
10.9
Voltage, frequency, and phase rotation
10.9.1
Phase rotation
Nominal system voltage range ............... 100 V
AC
550 V
AC
Nominal frequency range ....................... 14 Hz
500 Hz
Result displayed .................................... 1.2.3 or 3.2.1
10.9.2
Voltage
Measuring range (V)
Resolution (V)
Accuracy
0
550
1
(2 % of r 2 digits)
Result type ............................................ True r.m.s. (TRMS)
Nominal frequency range ....................... 0 Hz, 14 Hz
500 Hz
10.9.3
Frequency
Measuring range (Hz)
Resolution (Hz)
Accuracy
0.00
9.99
0.01
(0.2 % of r 1 digit)
10.0
499.9
0.1
Nominal voltage range ........................... 10 V
550 V
10.9.4
Online terminal voltage monitor
Measuring range (V)
Resolution (V)
Accuracy
10
550
1
(2 % of r 2 digits)