MI 3102(H) BT
Eurotest
XE
Measurements
42
5.4.3 Compensation of test leads resistance
This chapter describes how to compensate the test leads resistance in both continuity functions,
R LOWΩ and CONTINUITY. Compensation is required to eliminate the influence of test leads
resistance and the internal resistances of the instrument on the measured resistance. The lead
compensation is therefore a very important feature to obtain correct result.
symbol is displayed if the compensation was carried out successfully.
Circuits for compensating the resistance of test leads
Figure 5.16: Shorted test leads
Compensation of test leads resistance procedure
Select
R LOWΩ
or
CONTINUITY
function.
Connect
test cable to the instrument and short the test leads together (
Press the
TEST
key to perform resistance measurement.
Press the
CAL
key to compensate leads resistance.
Figure 5.17: Results with old calibration values
Figure 5.18: Results with new calibration
values
Note:
The highest value for lead compensation is 5
. If the resistance is higher the
compensation value is set back to default value.
is displayed if no calibration value is stored.