MI 3102(H) BT
Eurotest
XE
Technical specifications
102
10.14
Illumination
10.14.1
Illumination (Luxmeter sensor, type B)
Specified accuracy is valid for complete operating range.
Measuring range (lux)
Resolution (lux)
Accuracy
0.01
19.99
0.01
(5 % of r 2 digits)
20.0
199.9
0.1
(5 % of reading)
200
1999
1
2.00
19.99 k
10
Measurement principle ........................... silicon photodiode with V(
) filter
Spectral response error .......................... < 3.8 % according to CIE curve
Cosine error ............................................ < 2.5 % up to an incident angle of
85
O
Overall accuracy ..................................... matched to DIN 5032 class B standard
10.14.2
Illumination (Luxmeter sensor, type C)
Specified accuracy is valid for complete operating range.
Measuring range (lux)
Resolution (lux)
Accuracy
0.01
19.99
0.01
(10 % of r 3 digits)
20.0
199.9
0.1
(10 % of reading)
200
1999
1
2.00
19.99 k
10
Measurement principle ........................... silicon photodiode
Cosine error ............................................ < 2.5 % up to an incident angle of
85
O
Overall accuracy ..................................... matched to DIN 5032 class C standard