MI 3102(H) BT
Eurotest
XE
Technical specifications
95
10.7
PE conductor resistance
10.7.1
No RCD selected
PE conductor resistance
Measuring range (
)
Resolution (
)
Accuracy
0.00
19.99
0.01
(5 % of r 5 digits)
20.0
99.9
0.1
100.0
199.9
0,1
10 % of reading
200
1999
1
10.7.2
RCD selected
PE conductor resistance
Measuring range (
)
Resolution (
)
Accuracy
0.00
19.99
0.01
(5 % of r 10 digits)
20.0
99.9
0.1
100.0
199.9
0.1
10 % of reading
200
1999
1
Accuracy may be impaired in case of heavy noise on mains voltage.
Nominal voltage range ........................... 93 V
134 V (45 Hz
65 Hz)
185 V
266 V (45 Hz
65 Hz)
No trip out of RCD.