MI 3102(H) BT
Eurotest
XE
Technical specifications
90
10.2
Diagnostic test (MI 3102H BT only)
Dielectric absorption ratio DAR*
Measuring range
Resolution
Accuracy
0.01
9.99
0.01
(5% of r 2digits)
10.0
100.0
0.1
(5% of reading)
* Only for test voltages 500 V
DC
, 1000 V
DC
and 2500V
DC
. If any insulation resistance values
(
R
ISO
(15s) or R
ISO
(60s )) are over-ranged the
DAR
factor is not calculated. The result field is
blank: DAR:_____!
Polarization index PI**
Measuring range
Resolution
Accuracy
0.01
9.99
0.01
(5% of r 2digits)
10.0
100.0
0.1
(5% of reading)
** Only for test voltages 500 V
DC
, 1000 V
DC
and 2500V
DC
. If any insulation resistance values
(
R
ISO
(60s) or R
ISO
(10min)) are over-ranged the
PI
factor is not calculated. The result field is
blank: PI :_____!
10.3
Continuity
10.3.1
Resistance R LOW
Measuring range according to EN 61557 is 0.16
1999
.
Measuring range R (
)
Resolution (
)
Accuracy
0.00
19.99
0.01
(3 % of r 3 digits)
20.0
199.9
0.1
(5 % of reading)
200
1999
1
Measuring range R+, R- (
)
Resolution (
)
Accuracy
0.0
199.9
0.1
(5 % of r 5 digits)
200
1999
1
Open-circuit voltage ................................ 6.5 VDC
9 VDC
Measuring current ................................... min. 200 mA into load resistance of 2
Test lead compensation .......................... up to 5
The number of possible tests .................. > 2000, with a fully charged battery
Automatic polarity reversal of the test voltage.
10.3.2
Resistance CONTINUITY
Measuring range (
)
Resolution (
)
Accuracy
0.0
19.9
0.1
(5 % of r 3 digits)
20
1999
1
Open-circuit voltage ................................ 6.5 VDC
9 VDC
Short-circuit current ................................ max. 8.5 mA
Test lead compensation .......................... up to 5