MSI-3750 Digital Weight Indicator
◆
User Guide
Page 83
M E A S U R E M E N T S Y S T E M S I N T E R N A T I O N A L
Comm port while the weight is in motion.
INTERVAL: Selecting “ONCE” means that after the trigger condition, a single
transmission of data will occur. Selecting “ON” enables a timed transmission
rate. A print interval can be entered from 0 to 28,800 seconds. Used in conjunc-
tion with the Continuous output mode, this can generate a time based data-
logging system. An interval of 0 seconds means that the scale will output data at
the highest rate possible (limited to 4 readings per second). Interval used with
other control modes can be useful for control and monitoring purposes.
SETTINGS: The Settings menu is used to set standard Comm Port parameters:
BAUDRATE: Standard Baud Rates available are 300, 600, 1200, 2400, 4800,
9600, 19200. The 3750 defaults to 9600 Baud. See “To Set The Baud Rate”.
STOP BITS: Either 1 or 2 Stop Bits can be set. The 3750 Stop Bits default to 1.
DATA BITS: Either 7 or 8 Data Bits can be set. If 7 bit mode is set, Data
Characters above 127dec cannot be sent or received. The 3750 defaults to 8 bits.
PARITY: Parity mode is NONE, EVEN or ODD. The 3750 defaults to “NONE”,
parity off.
HANDSHAK(e): Comm Port Handshaking is set in this menu. Use XON/XOFF
for software handshaking commonly used in communicating with computers.
Use CTS/RTS for hardware handshaking. Use NONE for situations where no
handshaking is required or possible.
Summary of Contents for MSI-3750