MSI-3750 Digital Weight Indicator
◆
User Guide
DO NOT initiate this function unless you are prepared and
qualified to perform a complete initialization and Calibration
procedure.
CAUTION
1)
Expose the Calibration Seal Port by
removing the seal screw on the left side of
the 3750 Meter Casting with a Phillips
Head Screwdriver.
2)
Insert a small non-metallic screwdriver or
the stem side of a wooden “Q-Tip” and
press and release the switch button in the
hole followed immediately by the ZERO
key. The display reads “RESETCAL”.
3)
The first item in the Calibrate Setup Menu
appears Use the
UP
or
DOWN
scroll keys
to locate the next desired Calibrate opera-
tion.
➀
(see “Enable Calibration” for the location
of the Cal Port)
➁
ZERO
0
③
RESETCAL
③
STD CAL
This procedure erases all calibration constants, all the ID
Codes, all Set Points, all Print Strings, and resets all the
registers back to factory defaults. DO NOT initiate this function
unless you are prepared and qualified to perform a complete
initialization and Calibration procedure.
CAUTION
RESET CALIBRATION
Completely clears the internal EEROM of calibration settings, including capacity
and count-by (d). Does not erase other user functions such as ID Codes, or RS-
232 settings. Usually used for board replacement, changing capacity and/or
resolution, troubleshooting, or Load cell replacement. Does not change any
previously set Setup functions. A full calibration must follow this operation.
To Reset all the Calibration Constants
This procedure will not work if the scale is already in the CAL SETUP Menus. Push the EXIT
key to get out of “CAL SETUP”.
RESET ALL
Similar to the RESETCAL plus all setup functions are returned to defaults and all
ID Codes are erased. Starts the scale from scratch. Usually used for board
replacement, changing capacity and/or resolution, troubleshooting, or Load cell
replacement. A full calibration must follow this operation.