MSI-3750 Digital Weight Indicator
◆
User Guide
1)
Use the
UP
or
DOWN
scroll key to scroll
to the “AZM” message.
2)
Press
ENTER
to access the AZM submenu.
3)
Change AZM to “DISABLED” or “EN-
ABLED” as desired. using the
UP
or
DOWN
key.
4)
Push
ENTER
.
5)
If you selected “ENABLED” in the
previous step the “AZMRANGE” screen
will appear. Use the numeric keys to input
the weight range the system can zero out
automatically. Use “0” for standard AZM
operation. This sets the AZM window at
±
0.5d.
In this example, we’ll enter
±
1kg as the AZM
range.
Note: The AZM
±
range is limited to a maximum
of 255 `d`. If you exceed this range, the 3750
will return the error message “TOO BIG”.
6)
After keying in the desired AZM range,
push
ENTER
.
7)
Push
EXIT
to return to normal scale
operation, or push
UP
or
DOWN
to choose
another Calibrate Setup function.
➀
UV
F3
➠
UV
F3
➀
AZM
➁
ENTER
DISABLED
➂
UV
F3
ENABLED
√
ENTER
➄
SETUP
➄
1
ABC
1_
≈
ENTER
∆
EXIT
SETUP
!
Disabling Auto Zero Maintenance will degrade temperature
and drift performance of the MSI 3750. Also use caution in
setting the AZM window too large. This can cause the scale to
zero unexpectedly. MSI recommends keeping AZM below 5d
for most applications.
CAUTION
Enable/Disable AZM (Auto Zero Maintenance)
Note: In Legal-for-Trade systems the width of AZM can not be adjusted.