MSI-3750 Digital Weight Indicator
◆
User Guide
Page 113
M E A S U R E M E N T S Y S T E M S I N T E R N A T I O N A L
1)
Expose the Calibration
Seal Port by removing
the seal screw on the left
side of the 3750 Meter
Casting with a Phillips
Head Screwdriver.
Note: Dual Input 3750
Meters Only– Use the
ScaleID Key to select which
Scale (A or B) will be
calibrated. To Calibrate
both Scale Inputs, you must
complete the first. Exit out
of the Calibration Menus.
Select the other scale with
the ScaleID Key. Then
enable Calibration again.
2)
Insert a small non-
metallic screwdriver or
the wooden stem of a
“Q-Tip” and press the
switch button in the hole.
The display reads
“CALSETUP” for 2
seconds. Tip: A wooden
Golf Tee fits the hole
perfectly.
Dual Input (ABC) units
will also display the
message “LOAD EE?”
for 2 seconds.
3)
The first item in the
Calibrate Setup Menu
appears Use the
UP
or
DOWN
scroll keys to
locate the desired
Calibrate operation as
described in the follow-
ing procedures.
➀
Calibration
Seal Port
➁
SETUP
➂
STD CAL
To Enter the Calibration Menus