MSI-3750 Digital Weight Indicator
◆
User Guide
Page 73
M E A S U R E M E N T S Y S T E M S I N T E R N A T I O N A L
Figure 8
TEST SET POINT RELAYS
To test the connections and exercise the relays use the following procedure.
1)
Push
SETUP
followed immediately by the SET POINTS key (usually
F2
).
Push
ENTER
. You are now in the SET POINTS menu.
2)
Use the
DOWN
key to scroll to the “TEST SP” selection. Push
ENTER
.
3)
Use the
UP
key to step progressively through each set point. Each relay will
turn on in succession and the message display indicates which set point
relay is activated.
4)
When done testing the relays push
EXIT
twice to return to normal scale
operation.
Driving External Relays with Setpoint 8
External
Relay
+24V Typ
+24V
Return
10k
Ω
Resistor
Logic Level
N-Chan
Power MOSFET
D
G
S
TTL LO
GND
TTL HI
Setpoint 8
Isolated
Ground
Isolated
+5V
WARNING
This procedure will turn on and off any device connected to the relays.
Make sure that doing so will not cause a safety hazard or damage prop-
erty or goods.
WARNING
Potentially fatal voltages are present within the meter! Always remove
power from the meter and your circuit before opening the cabinet and
connecting the relay circuits. Ensure that the conductors to your circuit
are securely fastened in the terminal block and that the cable is securely
fastened in the watertight fitting. All wiring should comply with local
electrical codes.