FILTER
Use the LOW setting for most scale applications. It settles fastest and is intended
for general use. Use the MEDIUM setting when the scale is being used under
conditions that cause light to medium swinging. Use the HIGH setting when
there is a lot of scale motion or vibration. There is a time penalty to pay for using
the HIGH setting. The user should wait at least 5 seconds to ensure that the final
reading has settled (Motion indicator off).
Not available in some legal-for-trade systems or the FILTER menu is moved to the
Calibrate Setup Menu.
To Change the Filter Setting
➀
EXIT
SETUP
!
SETUP
2s delay
SETUP
➁
UV
F3
SETUP
➂
ENTER
SETUP
√
UV
F3
SETUP
➄
ENTER
SETUP
≈
EXIT
SETUP
!
1)
Push
SETUP
. Wait 2 seconds (or push
SETUP
again).The message reads “KEY
FCTN”.
2)
Push the
UP
scroll key to change the menu
selection to “FILTER”.
3)
Push
ENTER
.
4)
Select the “LOW”, “MEDIUM”, or
“HIGH” filter by using the
UP/DOWN
keys to scroll through the choices. In this
example, we’ll pick the MEDIUM filter
which is suitable to medium to high
vibration situations.
5)
Push
ENTER
.
6)
You are back in the
SETUP
menu. Return
to normal scale operation (MEDIUM Filter
on) by pushing
EXIT
(
SETUP
).