106
BA023001-en
3
After resetting the central module, the system must be switched off and on again.
Logging into the Service technician menu (see section "Operation > Settings >
Login Service technician menu") is then no longer possible online, but only via a
software key (maha-key).
If the central module has never been installed, a reset is not necessary and also not
possible.
Start screen of a new or reset board:
Service technician login is possible
only with offline activation
.
4
Perform a software update (see section "Operation > Service technician menu >
Software update").
Only a manual update can be performed, since no network setting is available yet.
5
After the software update has been completed, all further steps can now be
carried out in the same way as for commissioning.