Repair
9.1
Safety instructions
See section "Assembly > Safety instructions".
9.2
Configuring the control unit
In the spare parts case, the central module is not configured. Configuration is carried
out during initial commissioning as described below.
The following is required for replacing the control unit of the central module:
•
SN of the test bench
•
Replacement control (ZM)
•
Software key (installer key/token)
•
Software image
NOTICE
The technician key must not be older than the software image! It is recommended to
always download key and software at the same time. Both can be downloaded from
https://www.maha.de/de/software CONNECT DOWNLOADS.
The following steps must be observed when replacing the control unit:
1
Replace the central module with the main switch switched off.
2
Reset the central module to "Delivery status":
Resetting to the delivery status also deletes all error entries and the event logging
together with any software installed in the central module.