Operator’s Guide
Monthly Maintenance
Checking the Solution Temperatures
1. Touch:
•
Settings
•
Machine settings
•
PP temperature
2. Set the reference temperatures for the solutions and the dryer, if necessary:
DEV
40 °C ± 0.3
BX
38 °C ± 2
STB
37 °C ± 2
Dryer
65 °C
3. Calibrate the temperature sensors:
a. Touch
Readjust xxx
.
A window with a numeric keypad opens. The temperature of the selected
solution is indicated in the first line of the input field.
b. Remove the chemical filter.
Using a calibrated thermometer, measure the current temperature of the
processing solution through the free opening of the filter holder.
c. Enter the current temperature value via the numeric keypad.
The sensor will be calibrated after pressing
OK
.
NOTE: The
Reset reference
and
Reset calibration
buttons are used to enter the
standard values of the database.
8-22
KODAK PROFESSIONAL RP 30 Laser Printer
Summary of Contents for RP 30
Page 1: ...RP30 LASERPRINTER OPERATOR S GUIDE ...
Page 2: ...Operator s Guide KODAK PROFESSIONAL RP 30 Laser Printer P N 6B7444 December 2002 ...
Page 17: ...Operator s Guide 1 2 KODAK PROFESSIONAL RP 30 Laser Printer ...
Page 48: ...Overview Print Mode Menu KODAK PROFESSIONAL RP 30 Laser Printer 2 29 ...
Page 50: ...Overview Settings Menu with Machine Settings ar KODAK PROFESSIONAL RP 30 Laser Printer 2 31 ...
Page 51: ...Operator s Guide Test Menu 2 32 KODAK PROFESSIONAL RP 30 Laser Printer ...
Page 61: ...Operator s Guide 3 2 KODAK PROFESSIONAL RP 30 Laser Printer ...
Page 134: ...Operator s Guide 5 2 KODAK PROFESSIONAL RP 30 Laser Printer ...
Page 144: ...Operator s Guide 5 12 KODAK PROFESSIONAL RP 30 Laser Printer ...
Page 172: ...Operator s Guide 7 2 KODAK PROFESSIONAL RP 30 Laser Printer ...
Page 216: ...Operator s Guide 8 24 KODAK PROFESSIONAL RP 30 Laser Printer ...
Page 233: ......