7-28
TOS9213S/ TOS9213AS
IR:TIMER (I:TIM)
Sets the test time (TEST TIME) and ON/OFF of the timer function for an insulation
resistance test. Also inquires about the current test time and the ON/OFF status of
the timer function.
Message
Setting:
IR:TIMER_<
test time
, {ON|OFF}>
I:TIM_<
test time
, {ON|OFF}>
Query:
IR:TIMER?
I:TIM?
Program data
<test time>
Data format: Real number
Set value:
0.5 to 999
Resolution:
0.1 for 0.5 to 99.9, 1 for 100 to 999
Unit:
s
Program data
<{ON|OFF}>
Data format: Character
Set value:
OFF (0), ON (1)
(Example) To set the test time to 5 s,
I:TIM 5,ON
Response
To I:TIM
?
, the current test time and the ON/OFF status of the timer function are
returned.
(Example) If the test time is currently set to 2 s and the timer function to OFF,
“
2.0,0
” is returned.
IR:RISETIME (I:RTIM)
Sets the voltage rise time (RISE TIME) for an insulation resistance test. Also
inquires about the set or setting voltage rise time.
Message
Setting:
IR:RISETIME_<
voltage rise time
>
I:RTIM_<
voltage rise time
>
Query:
IR:RISETIME?
I:RTIM?
Program data
Data format: Real number
Set value:
0.1 to 200
Resolution:
0.1 for 0.1 to 99.9, 1 for 100 to 200
Unit:
s
(Example) To set the voltage rise time to 1 s,
I:RTIM 1
Response
To I:RTIM
?
, the current voltage rise time is returned.
(Example) If the voltage rise time is currently set to 2.0 s,
2.0
is returned.