5-54
TOS9213S/ TOS9213AS
Setting the test conditions (Memory No.) and interval time
• If a test condition under the preset memory number is set to TIMER OFF, a pro-
gram test ends once the STOP switch is pressed, preventing the next test from
being performed.
For each step, set the memory number of the test conditions to be used and the inter-
val between steps. The program executes the test in order of step number.
When the cursor is positioned at the step number, the screen can be scrolled using
the rotary knob. The
keys can be used to perform scrolling, regardless of the
location of the cursor.
Upon completion of each step, “END” or “RET” is displayed.
1.
To mo
v
e the c
u
rsor to the step n
u
m
b
er, press the SHIFT + F1 keys.
2.
Mo
v
e the c
u
rsor to the step
b
elo
w
the step to
b
e inserted. To add a step
to the pre
v
io
u
s step, mo
v
e the c
u
rsor to E
N
D or RET.
3.
Press the F1 key (I
N
S). A step (DC
W
00-0.2s)
w
ill
b
e inserted at the
c
u
rsor.
4.
Using the key, mo
v
e the c
u
rsor to DC
W
to the right of the step n
u
m-
b
er.
5.
Using the rotary kno
b
, set the target test (DC
W
, IR, EC).
To control the earth continuity tester TOS6200, select EC. See "Chapter 8 Con-
trolling the TOS6200".
6.
Using the key, mo
v
e the c
u
rsor to the memory n
u
m
b
er.
7.
Using the rotary kno
b
, set a memory n
u
m
b
er.
8.
Using the key, mo
v
e the c
u
rsor to the inter
v
al time to the right of the
memory n
u
m
b
er.
9.
Using the rotary kno
b
, set an inter
v
al time (from 0.2 s to 9.9 s, HOLD).
If the interval time is set to HOLD, the next step starts when the START switch
is pressed with the specified step in the HOLD status.
To delete a step, move the cursor to the corresponding step number and press the F2
key (DEL).
To change the interval time or the memory number of a step, move the cursor to the
target and make changes using the rotary knob.
NOTE